HC2A-S Humidity probe
Reliable humidity measurement in open air with the HygroClip2 ADVANCED
The HC2A-S humidity probe provides precise measurements with an accuracy of ±0.8%RH and ±0.1°C. Ideal for applications that require reliable humidity monitoring.
Specifications:
- Humidity: 0-100 %RH
- Temperature: -50 to +100 °C
- HYGROMER®HT-1 sensor for accurate measurement results
- Long-term stable with < 1 %RH/year

SPECIFICATION HC2A-S Humidity Probe
Measuring range | |
---|---|
humidity | 0…100 %rF |
Temperature | -50... 100 °C |
Measurement inaccuracy | |
---|---|
humidity | ±0.8 %RH at 10...30 °C |
humidity | ±1.3 %RH at -10…+10 °C and 30…50 °C |
humidity | ±1.8 %RH at 50…70 °C |
humidity | ±2.8 %RH at 70…100 °C |
humidity | ±2.3 %RH at -30…-10 °C |
humidity | ±3.3 %RH at -50…-30 °C |
Temperature | ±0.1 °C at 10...30 °C |
Description
The HC2A-S is the preferred choice for indoor humidity measurement. Its advanced design and high accuracy make it ideal for various applications. It can be connected to the multimeter with an extra cable
Swema 3000/md/mdH connect.